Safety and efficacy of cerebral embolic protection in transcatheter aortic valve implantation: an updated meta-analysis

AsiaIntervention 2024;10:51-59 | 10.4244/AIJ-D-23-00022

Nicholas Tan1, MBBS; Gao Fei1, PhD; Mohammed Rizwan Amanullah1, MBBS; Soo Teik Lim1,2, MBBS; Zameer Abdul Aziz3, MBBS; Sivaraj Govindasamy3, MBBS; Victor Tar Toong Chao3, MBBS, FRCS; See Hooi Ewe1,2, MBBS, PhD; Kay Woon Ho1,2, MBBS; Jonathan Yap1,2, MBBS, MPH

1. Department of Cardiology, National Heart Centre Singapore, Singapore; 2. Duke-NUS Medical School, Singapore; 3. Department of Cardiothoracic Surgery, National Heart Centre Singapore, Singapore

Abstract

Background: The use of cerebral embolic protection devices during transcatheter aortic valve implantation (TAVI) reveals conflicting data.

Aims: This updated meta-analysis aims to evaluate the efficacy and safety of the SENTINEL Cerebral Protection System.

Methods: A literature search for relevant studies up to September 2022 was performed. Study outcomes were divided based on time period – overall (up to 30 days) and short (≤7 days). The outcomes studied include stroke (disabling, non-disabling), mortality, neuroimaging findings, transient ischaemic attack, acute kidney injury and major vascular and bleeding complications.

Results: A total of 15 studies involving 294,134 patients were included. Regarding overall outcomes, significant reductions were noted for mortality (odds ratio [OR] 0.60, 95% confidence interval [CI]: 0.41-0.88; p=0.008), all stroke (OR 0.64, 95% CI: 0.46-0.88; p=0.006) and disabling stroke (OR 0.42, 95% CI: 0.23-0.74; p=0.003) using the SENTINEL device. No significant differences were noted for other outcomes. There was significant heterogeneity across the studies for mortality (p=0.013) and all stroke (p=0.003). Including only randomised data (n=4), there was only significant reduction in the incidence of disabling stroke (OR 0.39, 95% CI: 0.17-0.89; p=0.026) in the SENTINEL group. In studies reporting ≤7-day outcomes (n=8), use of the SENTINEL device demonstrated significantly lower rates of all stroke (p<0.001), disabling stroke (p<0.001) and major bleeding complications (p=0.02). No differences in neuroimaging outcomes were noted.

Conclusions: In this updated meta-analysis, use of the SENTINEL Cerebral Protection System was associated with lower rates of mortality, all stroke and disabling stroke, although significant heterogeneity was noted for mortality and all stroke. Including exclusively randomised data, there was only significant reduction in the incidence of disabling stroke. No significant adverse outcomes with device use were noted.

Abbreviations

  • CPS: cerebral protection system
  • TAVI: transcatheter aortic valve implantation
  • EPD: embolic protection device

Introduction

Stroke during transcatheter aortic valve implantation (TAVI) is not an uncommon complication12, with an occurrence of about 2%3. This often results in significant mortality and morbidity4567. With the extension of TAVI to low-risk patients, stroke prevention takes on increasing importance, as this could adversely affect the physical and cognitive function in a subgroup with a longer life expectancy89. To potentially minimise the risk of strokes, embolic protection devices (EPDs) were designed to help capture embolic debris during TAVI.

The only EPD approved for use by the U.S. Food and Drug Administration to date is the SENTINEL Cerebral Protection System (CPS; Boston Scientific)10. The device comprises two filters, with the proximal filter positioned in the brachiocephalic trunk and the distal filter in the left common carotid artery.

Conflicting data on the efficacy of the SENTINEL device exist in the results of randomised controlled trials (RCTs), although these trials were limited by their small sample sizes111213. A meta-analysis on the subject14 reported a significantly lower risk of stroke, mortality and major bleeding. However, this did not include the recently published PROTECTED TAVR trial, which is the largest RCT to date evaluating the SENTINEL CPS15. The PROTECTED TAVR trial, comprising 3,000 patients, showed no significant reduction in clinical stroke after using the SENTINEL device. However, there was a significant reduction in the secondary endpoint of disabling stroke. This updated meta-analysis, which includes the PROTECTED TAVR trial, aims to evaluate the clinical efficacy and safety of the SENTINEL CPS.

Methods

LITERATURE SEARCH

This meta-analysis was written in accordance with the Preferred Reporting Items for Systematic reviews and Meta-Analysis (PRISMA) guidelines (Supplementary Appendix 1). A comprehensive literature search using the Cochrane Library and PubMed databases was performed for relevant articles up to September 2022. The following Medical Subject Heading (MeSH) words were used for the search: “Embolic Protection Devices”, “SENTINEL Cerebral Protection System”, “Transcatheter Aortic Valve Implantation”, and “Transcatheter Aortic Valve Replacement”. The bibliographies of relevant studies were also hand-searched to find more studies which met our inclusion criteria. There were no language restrictions applied.

STUDY SELECTION AND OUTCOMES

Publications were reviewed independently by one author (N. Tan) and were included if they reported at least 1 of the following clinical outcomes: all stroke, disabling stroke, non-disabling stroke, mortality, transient ischaemic attack (TIA), acute kidney injury (AKI), major bleeding complications, and major vascular complications. The primary focus was on ischaemic strokes − for which the SENTINEL device is primarily designed to prevent − and thus, haemorrhagic strokes, if reported, were excluded. Some studies, however, did not differentiate between ischaemic and haemorrhagic strokes, and we were unable to exclude the latter in these cases. Other inclusion criteria were studies reporting brain magnetic resonance imaging (MRI) radiological outcomes with respect to ischaemic cerebral lesions post-TAVI. Exclusion criteria included single-arm studies (which only evaluated the efficacy of TAVI with the SENTINEL CPS), studies where other EPDs were included and studies which duplicated databases (for which we included the study with the larger patient pool). This selection process is visually depicted in Supplementary Figure 1.

DATA EXTRACTION AND QUALITY ASSESSMENT

Conflicts regarding the inclusion of studies were discussed and resolved with a second author (J. Yap). The risk of bias was assessed using the Cochrane Collaboration’s tool for assessing risk of bias in randomised controlled trials (RoB 2.0)16 and Risk of Bias in Non-randomized Studies – of Interventions (ROBINS-I) tool17. These results are shown in Supplementary Table 1.

STATISTICAL ANALYSIS

Clinical outcomes were analysed based on timepoints – overall outcomes (up to 30 days) and short-term outcomes (≤7 days). Studies which only provided ≤7-day outcomes were included in the eventual overall outcomes as well. If a study provided both >7-day and ≤7-day outcomes, these were included under both overall and short-term outcomes, respectively. The stroke rates at ≤7 days may portray a more accurate picture of periprocedural outcomes18, as other factors (such as atrial fibrillation) may confound complication rates at the 30-day mark. Thirty-day complication rates help to capture delayed complications for better safety assessment. We utilised the formulae suggested by Wan et al19 to estimate the mean and standard deviation if the study only provided the median and interquartile range for that outcome. We estimated the standardised mean difference (SMD) and odds ratio (OR) via the random-effects model using the Mantel-Haenszel method, with 95% confidence intervals (CI), for radiological and clinical outcomes, respectively. Heterogeneity between studies was evaluated using the Q statistic and I2 test. A sensitivity analysis including only the RCTs was performed for overall outcomes; there were insufficient data to do so for short-term outcomes.

The presence of publication bias was assessed by visually examining the degree of symmetry in Begg’s funnel plot (Supplementary Figure 2) between treatment effects and their standard error (SE) as well as with Egger’s test. Funnel plots were constructed when there were at least 5 studies assessing the association of the SENTINEL device with a particular outcome. A vertical line indicates the estimate based on the model. A pseudo confidence interval region is drawn around this value with bounds equal to ±1.96 times the SE. The studies outside of the pseudo confidence interval region are labelled. Funnel plots may show asymmetry in the absence of publication bias20. To investigate whether heterogeneity may be further explained by differences in characteristics of the studies, we performed univariate regression on age, sex and Society of Thoracic Surgeons (STS) score. Data were extracted and analysed by another independent author (G. Fei) using R with the metafor package (R Foundation for Statistical Computing)21 for meta-analysis.

Results

A total of 15 publications involving 294,134 patients (25,910 with the SENTINEL device and 268,224 without the SENTINEL device) were included. These comprised 4 randomised controlled trials, 8 propensity-matched and 3 cohort studies. The mean age was 80.6 years old, the mean proportion of males was 50.7%, and the mean STS score was 5.27 (Table 1).

Table 1. Summary of included studies.

Author,year Typeof study SENTINELarm (n) No SENTINEL arm (n) Outcomes studied Timing of follow-up Age (years) Male (%) STS score Type of valve
Haussig et al, 2016# 12 Randomised controlled trial 50 50 Mortality, all stroke, disabling stroke, non-disabling stroke, AKI, major vascular complications, major bleeding complicationsMedian total new lesion volume, median number of new lesions 2 days,
7 days,
30 days*2 days, 7 days for neuroimaging
80.0±5.1 (SENTINEL)79.3±4.1 (No SENTINEL) 43.0% (43/100) 5.6±3.2 (SENTINEL)5.2±2.7 (No SENTINEL) Medtronic CoreValve
Kapadia et al, 201713 Randomised controlled trial 234 (121 in device arm, 123 in safety arm) 119 Mortality, all stroke, disabling stroke, non-disabling stroke, TIA, AKI, major vascular complicationsMedian total new lesion volume, median number of new lesions 30 days2-7 days for neuroimaging 83.4 (78.0-88.2) 47.9% (174/363) 6.0 (4.2-8.1) Medtronic CoreValve/Evolut R, Edwards SAPIEN XT/3
Kapadia et al, 202215 Randomised controlled trial 1,501 1,499 Mortality, all stroke, disabling stroke, non-disabling stroke, TIA, AKI, major vascular complications, major bleeding complications 3 days or in-hospital (whichever  earliest) 78.9±8.0 (SENTINEL)78.9±7.8 (No SENTINEL) 60.1%(1,803/3,000) 3.3±2.7 (SENTINEL)3.4±2.8 (No SENTINEL) Medtronic Evolut R/Evolut PRO, Edwards SAPIEN 3, Lotus, ACURATE, Portico
Van Mieghem et al, 201611 Randomised controlled trial 32 33 Mortality, all stroke, disabling stroke, non-disabling stroke, AKI, major vascular complications, major bleeding complicationsMedian total new lesion volume, single lesion volume, median number of new lesions <24 hours,
30 days*,
6 months5 days for neuroimaging
82 (78-85) 52.0% (34/65) 4.8 (3.4-7.2) Medtronic CoreValve, Edwards SAPIEN XT/3, Portico
Alkhouli et al, 2020§ 31 Propensity-matched 2,732 8,253 Mortality, all stroke, disabling stroke In-hospital 80.0±9.2 (SENTINEL)80.1±9.2 (No SENTINEL) 58.5% (SENTINEL)55.7% (No SENTINEL) – –
Butala et al, 202132 Propensity-matched 12,409 110,777 Mortality, all stroke, major bleeding complications 30 days*, in-hospital 79.0±8.9 (SENTINEL)79.4±8.8 (No SENTINEL) 55.1% (67,920/123,186) – Medtronic CoreValve, Edwards SAPIEN
Isogai et al, 202237 Propensity-matched 1,802 1,037 Mortality, all stroke, TIA 3 days or in-hospital (whichever earliest) 79.2±9.5 58.5% 4.5 (2.9-7.0) (SENTINEL)6.0 (4.2-8.6) (No SENTINEL) Medtronic CoreValve/Evolut R/Evolut PRO/Evolut PRO+, Edwards SAPIEN XT/3/3 Ultra
Kroon et al, 201933 Propensity-matched 333 333 Mortality, all stroke, disabling stroke, non-disabling stroke, TIA, AKI, major vascular complications, major bleeding complications Procedural,
24 hours,
30 days*, in-hospital
81 (76-85) 49.2% (328/666) 4.4 (3.0-6.6) Medtronic CoreValve/Evolut R, Edwards SAPIEN XT/3, Lotus, Portico
Khan et al, 2021¶ 34 Propensity-matched 4,380 103,935 Mortality, all stroke, TIA, AKI, major vascular complications, major bleeding complications In-hospital 81 (74-87) (SENTINEL)81 (75-86) (No SENTINEL) 53.7% (58,160/108,315) – –
Seeger et al, 201718 Propensity-matched 280 280 Mortality, all stroke, disabling stroke, non-disabling stroke, AKI, major vascular complications, major bleeding complications 7 days 80.6±6.0 (SENTINEL)80.9±6.4 (No SENTINEL) 54.6% (306/560) 6.2±4.2 (SENTINEL)6.9±5.0 (No SENTINEL) Medtronic Evolut R, Edwards SAPIEN 3, Lotus
Seeger et al, 201935 Propensity-matched 533 533 All stroke, disabling stroke, non-disabling stroke 3 days 81.2±7.1 (SENTINEL)81.0±6.6 (No SENTINEL) 47.1% (502/1,066) 6.4±4.2 (SENTINEL)6.6±4.9 (No SENTINEL) –
Stachon et al, 2021## 30 Propensity-matched 1,564 40,090 Mortality, all stroke, AKI In-hospital 80.6±6.4 (SENTINEL)81.1±6.0 (No SENTINEL) 46.2% (SENTINEL)46.8% (No SENTINEL) – –
Dona et al, 2022 ‡36 Cohort 213 198 Mortality, all stroke, disabling stroke, non-disabling stroke, TIA, AKI 3 days*,
12 months 
80.4±6.7 52.6% (216/411) – Medtronic Evolut R, Portico, Symetis ACURATE, Centera Valve, Allegra Valve NVT
Seeger et al, 202038 Cohort 92 904 All stroke, disabling stroke 3 days,
30 days*, in-hospital
80.8 49.2% 6.0±6.9 Lotus
Voss et al, 201839 Cohort 39 352 AKI, major vascular complications In-hospital 79.1±7.3 (SENTINEL)79.5±7.2 (No SENTINEL) 50.9% (199/391) 3.5±2.1 (SENTINEL)4.4±4.2 (No SENTINEL) Medtronic CoreValve/Evolut R, Edwards SAPIEN 3
Values are presented as mean±standard deviation or median (interquartile range). #In the CLEAN-TAVI trial by Haussig et al, 2 day and 7-day outcomes were not used, as 3-day outcomes by Seeger et al 2019 were used instead (The latter study pooled patients from CLEAN-TAVI, SENTINEL-Ulm, SENTINEL US IDE trial). Two-day timepoints for neuroimaging were also used over the 7-day timepoints. §In this study, although it was not explicitly stated that all devices used were SENTINEL, it was presumed so, as patients in the device group were selected between 1 July 2017 and 31 December 2018, when the SENTINEL was the only approved EPD in the US market. ¶In this study, it was not explicitly stated that all devices used were SENTINEL, but it was presumed so, as the study was based in the USA where the SENTINEL is the only FDA-approved device. ##In this study, although it was not explicitly stated that all devices used were SENTINEL, but it was presumed that majority was SENTINEL as the SENTINEL received approval in Europe in 2014 and patient data collected was from 2015 to 2017. ‡Only the types of self-expanding valves were elaborated on, which comprised 64.0% (263/411) of the patient pool. There was no mention of balloon-expandable valves. Other notes: *In studies with multiple timepoints, the * indicates the timepoint which was used in our meta-analysis for overall outcomes. For male %, separate values for the SENTINEL and no SENTINEL in % were provided if study did not provide the exact number of patients to calculate an overall average male % between the 2 groups. AKI: acute kidney injury; EPD: embolic protection device; FDA: U.S. Food and Drug Administration; STS: Society of Thoracic Surgeons; TIA: transient ischaemic attack

OVERALL OUTCOMES

Regarding overall outcomes (Figure 1), significant reductions were noted for mortality (OR 0.60, 95% CI: 0.41-0.88; p=0.008), all stroke (OR 0.64, 95% CI: 0.46-0.88; p=0.006) and disabling stroke (OR 0.42, 95% CI: 0.23-0.74; p=0.003) in the SENTINEL group. No significant differences were reported for non-disabling stroke (p=0.46), TIA (p=0.26), AKI (p=0.19), major bleeding complications (p=0.11) or major vascular complications (p=0.41). There was significant heterogeneity across the studies for mortality (I2=69.7%; p=0.013) and all stroke (I2=75.0%; p=0.003). A sensitivity analysis (Supplementary Figure 3) was performed including only the 4 RCTs. There were no differences in mortality (OR 0.95, 95% CI: 0.35-2.61; p=0.92) or all stroke (OR 0.74, 95% CI: 0.50-1.10; p=0.14) when comparing the SENTINEL versus no SENTINEL groups. However, there was a significant reduction in the rate of disabling stroke (OR 0.39, 95% CI: 0.17-0.89; p=0.026) in the SENTINEL group. The visual examination of the funnel plots and the results of Egger’s test suggested no publication bias for overall mortality (p=0.83) and low publication bias for all stroke (p=0.049).

Figure 1. Forest plots of overall outcomes. CI: confidence interval; EPD: embolic protection device; RE: random effects

SHORT-TERM OUTCOMES

For the studies reporting ≤7-day outcomes (n=8) (Figure 2), significant reductions were noted for all stroke (OR 0.42, 95% CI: 0.27-0.65; p<0.001), disabling stroke (OR 0.28, 95% CI: 0.15-0.50; p<0.001) and major bleeding complications (OR 0.29, 95% CI: 0.10-0.79; p=0.02) in the SENTINEL group. No significant differences were reported for mortality (p=0.50), non-disabling stroke (p=0.20), major vascular complications (p=0.46) or AKI (p=1). There was no significant heterogeneity across all the outcomes (p>0.05).

Figure 2. Forest plots of short-term outcomes. CI: confidence interval; EPD: embolic protection device; RE: random effects

NEUROIMAGING OUTCOMES

For neuroimaging outcomes (n=3) (Figure 3), no significant differences in total new lesion volume (SMD −0.48, 95% CI: −1.00 to 0.04; p=0.07), number of new lesions (SMD −0.52, 95% CI: −1.35 to 0.31; p=0.22) or patients with new lesions (OR 0.54, 95% CI: 0.12-2.40; p=0.42) were seen in the SENTINEL group compared to the control group. Significant heterogeneity was noted for total new lesion volume (I2=76.3%; p=0.006) and number of new lesions (I2=90.4%; p=0.001).

Figure 3. Forest plots of neuroimaging outcomes. CI: confidence interval; EPD: embolic protection device; RE: random effects; SD: standard deviation; SMD: standardised mean difference

Discussion

In this large, updated meta-analysis including the recently published PROTECTED TAVR trial, the important findings are as follows: 1) Cerebral embolic protection with the SENTINEL device was associated with significantly lower rates of mortality, overall stroke and disabling stroke, with short-term (≤7-day) data showing a similar reduction in overall and disabling stroke and randomised data primarily showing a reduction in disabling stroke. 2) There were no significant differences in neuroimaging outcomes (albeit these data were only available in a smaller number of studies). 3) There were no significant differences in TIA, AKI, major bleeding complications or major vascular complications.

EFFICACY

Strokes tend to occur periprocedurally72223, with 50% occurring within 2 days. In this study, significant reductions in the incidence of all stroke and disabling stroke were noted. This reduction in stroke rate has been corroborated in several meta-analyses on the topic. Radwan et al24 showed a reduction in stroke (OR 0.82, 95% CI: 0.71-0.94; p=0.004) with use of the SENTINEL device. Similarly, Ndunda et al14 showed significant reductions in all stroke (OR 0.51, 95% CI: 0.29-0.90; p=0.02) at 30 days. This may potentially be attributed to the mechanism of the device itself, which involves capturing embolic debris during the procedure and, subsequently, lowering the rates of ischaemic stroke within the periprocedural period. Including only randomised data, there was a reduction in the rate of disabling stroke but not for all stroke. The PROTECTED TAVR trial15 randomised 3,000 patients and showed no significant difference in the occurence of clinical stroke between the SENTINEL and the control groups (2.3% vs 2.9%; p=0.30). While the trial was not powered to evaluate disabling stroke, results showed that it occurred in significantly fewer patients in the SENTINEL group compared to the control group (0.5% vs 1.3%; p=0.02).

In addition, we observed a significant reduction in mortality; this was also reported by both Radwan et al24 and Ndunda et al14. However, numerous studies have noted a significant increase in mortality and morbidity in patients with stroke post-TAVI62526. This reduction in mortality with the device may, in part, be from a reduction in the stroke rate, especially disabling strokes. Of note, when only randomised data were included, this impact on mortality was attenuated. The PROTECTED TAVR trial did not show any difference in mortality between the cerebral embolic protection (CEP) and the control groups (0.5% vs 0.3%). Whether the increased sample size and power in meta-analysis are required to show such a mortality effect, or if the confounders not addressed in non-randomised data led to such results will be the work for future larger-scale randomised trials to address.

Neuroimaging may provide additional information, as new subclinical ischaemic lesions after TAVI are seen in approximately 60% to 90% of patients on a brain MRI1227. New lesions are often seen in both hemispheres and are multiple8 – further suggestive of an embolic source. There is a discrepancy between clinically apparent strokes and the incidence of new ischaemic lesions seen on MRI post-TAVI, suggesting that some lesions might be silent. These lesions may affect long-term cognitive status, as studies have shown that silent lesions may cause a more pronounced cognitive decline and an increased risk of dementia2829. This meta-analysis did not show a significant reduction in total new lesion volume, number of new lesions or patients with new lesions on neuroimaging. The reasons for these findings could be multifactorial, as cerebral embolisation during TAVI is still possible even with the SENTINEL device. First, the SENTINEL CPS only has filters covering the brachiocephalic trunk and left common carotid artery, thus, leaving the left vertebral artery unprotected and susceptible to embolism. Second, manipulation of the SENTINEL device within atherosclerotic arteries may cause embolism directly. Thrombi may also form on the surface of the protection device itself during the earlier phases of the procedure when full anticoagulation may not yet be in effect30. Third, the SENTINEL filter only has 1 size, and malapposition may occur, leading to incomplete sealing of arteries and allowing embolic debris to slip by. Fourth, embolic debris may be smaller than the pore size of the filters (140 μm in diameter), thus, it can pass through the filter. Lastly, the studies included in this meta-analysis also had several issues including incomplete MRI follow-ups and small numbers.

SAFETY

Use of the SENTINEL system was shown to be safe, with no significant differences in safety outcomes such as AKI, or major vascular or bleeding complications. The safety profile of the device has been corroborated by other meta-analyses, with both Radwan et al24 and Ndunda et al14 showing no significant differences for AKI or major vascular complications at 30 days. The latest PROTECTED TAVR trial15 also showed no major differences between the SENTINEL and the control groups in terms of AKI (0.5% vs 0.5%), and only 1 (0.1%) patient suffered from a vascular complication due to the insertion of the SENTINEL device.

FUTURE DIRECTIONS

The ongoing British Heart Foundation (BHF) PROTECT-TAVI, which is a randomised trial aiming to recruit 7,730 patients, will be assessing the clinical benefits of the SENTINEL device, not only at short-term (72 hours) intervals, but also at longer (12 months) intervals as well. This will provide further and larger-scale data on the safety and efficacy of cerebral embolic protection devices. Beyond randomised clinical trials on the topic, another area of clinical interest is how to identify patients who may benefit from CEP. Thus far, data on this have been mixed, and there is no established scoring system which is routinely deployed in clinical practice to aid in the identification of such patients. This will be the work of future research.

Limitations

There are several limitations of this study. First, the impact of using different types of valves (e.g., balloon-expandable vs self-expanding valves) and procedural techniques (e.g., predilation of the aortic valve, transfemoral approach vs alternative approaches) on outcomes was unable to be accounted for. Second, there was variation in the timing of outcomes assessed across the studies. Third, the numbers, as well as sample sizes, for studies which looked at neuroimaging outcomes were small, which rendered them underpowered. Fourth, longer-term cognitive outcomes were not available, and these may be important, as silent infarcts have been linked with neurocognitive decline2829. Lastly, the lack of patient-level data precluded the identification of certain patient subsets who may benefit from CEP. This would be best elucidated with a patient-level data meta-analysis.

Conclusions

In this updated meta-analysis, use of the SENTINEL system was associated with lower rates of mortality, all stroke and disabling stroke, although significant heterogeneity was noted for the studies reporting mortality and all stroke. Including exclusively randomised data, there was only a significant reduction in disabling stroke. No significant adverse outcomes with device use were observed.

Impact on daily practice

Stroke during TAVI is not uncommon, thus, emphasising the importance of its prevention. Our paper, which evaluates the clinical efficacy and safety of the SENTINEL cerebral protection system during TAVI, found that it was associated with lower mortality, all stroke and disabling stroke. We hope these data will provide guidance to TAVI operators on the use of the SENTINEL system.

Conflict of interest statement

J. Yap received speaker honoraria from Biosensors, Biotronik, Boston Scientific, Edwards Lifesciences, Johnson & Johnson, Kaneka, Medtronic, and Terumo. K.W. Ho received speaker fees from Edwards Lifesciences, Medtronic, and Abbott Medical. S.H. Ewe received speaker fees from Abbott Medical, Philips, and GE HealthCare. The other authors have no conflicts of interest to declare.

Share

Supplementary data

Supplementary Appendix 1. PRISMA checklist. Supplementary Table 1. Risk-of-bias assessment. Supplementary Figure 1. Flowchart of study selection. Supplementary Figure 2. Funnel plot of outcomes. Supplementary Figure 3. Forest plots of overall outcomes including only randomised data.

References

  • Ghanem A, Müller A, Nähle CP, Kocurek J, Werner N, Hammerstingl C, Schild HH, Schwab JO, Mellert F, Fimmers R, Nickenig G, Thomas D. Risk and fate of cerebral embolism after transfemoral aortic valve implantation: a prospective pilot study with diffusion-weighted magnetic resonance imaging. J Am Coll Cardiol 2010;55:1427-32
  • Arnold M, Schulz-Heise S, Achenbach S, Ott S, Dörfler A, Ropers D, Feyrer R, Einhaus F, Loders S, Mahmoud F, Roerick O, Daniel WG, Weyand M, Ensminger SM, Ludwig J. Embolic cerebral insults after transapical aortic valve implantation detected by magnetic resonance imaging. JACC Cardiovasc Interv 2010;3:1126-32
  • Stachon P, Kaier K, Zirlik A, Bothe W, Heidt T, Zehender M, Bode C, von Zur Mühlen C. Risk-Adjusted Comparison of In-Hospital Outcomes of Transcatheter and Surgical Aortic Valve Replacement. J Am Heart Assoc 2019;8:e011504
  • Huded CP, Tuzcu EM, Krishnaswamy A, Mick SL, Kleiman NS, Svensson LG, Carroll J, Thourani VH, Kirtane AJ, Manandhar P, Kosinski AS, Vemulapalli S, Kapadia SR. Association Between Transcatheter Aortic Valve Replacement and Early Postprocedural Stroke. JAMA 2019;321:2306-15
  • Kapadia SR, Huded CP, Kodali SK, Svensson LG, Tuzcu EM, Baron SJ, Cohen DJ, Miller DC, Thourani VH, Herrmann HC, Mack MJ, Szerlip M, Makkar RR, Webb JG, Smith CR, Rajeswaran J, Blackstone EH, Leon MB; PARTNER Trial Investigators. Stroke After Surgical Versus Transfemoral Transcatheter Aortic Valve Replacement in the PARTNER Trial. J Am Coll Cardiol 2018;72:2415-26
  • Vlastra W, Jimenez-Quevedo P, Tchétché D, Chandrasekhar J, de Brito FS Jr, Barbanti M, Kornowski R, Latib A, D’Onofrio A, Ribichini F, Baan J, Tijssen JGP, De la Torre Hernandez JM, Dumonteil N, Sarmento-Leite R, Sartori S, Rosato S, Tarantini G, Lunardi M, Orvin K, Pagnesi M, Hernandez-Antolin R, Modine T, Dangas G, Mehran R, Piek JJ, Delewi R. Predictors, Incidence, and Outcomes of Patients Undergoing Transfemoral Transcatheter Aortic Valve Implantation Complicated by Stroke. Circ Cardiovasc Interv 2019;12:e007546
  • Kapadia S, Agarwal S, Miller DC, Webb JG, Mack M, Ellis S, Herrmann HC, Pichard AD, Tuzcu EM, Svensson LG, Smith CR, Rajeswaran J, Ehrlinger J, Kodali S, Makkar R, Thourani VH, Blackstone EH, Leon MB. Insights Into Timing, Risk Factors, and Outcomes of Stroke and Transient Ischemic Attack After Transcatheter Aortic Valve Replacement in the PARTNER Trial (Placement of Aortic Transcatheter Valves). Circ Cardiovasc Interv 2016;9:e002981
  • Mastoris I, Schoos MM, Dangas GD, Mehran R. Stroke after transcatheter aortic valve replacement: incidence, risk factors, prognosis, and preventive strategies. Clin Cardiol 2014;37:756-64
  • Pagnesi M, Martino EA, Chiarito M, Mangieri A, Jabbour RJ, Van Mieghem NM, Kodali SK, Godino C, Landoni G, Colombo A, Latib A. Silent cerebral injury after transcatheter aortic valve implantation and the preventive role of embolic protection devices: A systematic review and meta-analysis. Int J Cardiol 2016;221:97-106
  • Rogers T, Alraies MC, Torguson R, Waksman R. Overview of the 2017 US Food and Drug Administration Circulatory System Devices Panel meeting on the Sentinel Cerebral Protection System. Am Heart J 2017;192:113-9
  • Van Mieghem NM, van Gils L, Ahmad H, van Kesteren F, van der Werf HW, Brueren G, Storm M, Lenzen M, Daemen J, van den Heuvel AF, Tonino P, Baan J, Koudstaal PJ, Schipper ME, van der Lugt A, de Jaegere PP. Filter-based cerebral embolic protection with transcatheter aortic valve implantation: the randomised MISTRAL-C trial. EuroIntervention 2016;12:499-507
  • Haussig S, Mangner N, Dwyer MG, Lehmkuhl L, Lücke C, Woitek F, Holzhey DM, Mohr FW, Gutberlet M, Zivadinov R, Schuler G, Linke A. Effect of a Cerebral Protection Device on Brain Lesions Following Transcatheter Aortic Valve Implantation in Patients With Severe Aortic Stenosis: The CLEAN-TAVI Randomized Clinical Trial. JAMA 2016;316:592-601
  • Kapadia SR, Kodali S, Makkar R, Mehran R, Lazar RM, Zivadinov R, Dwyer MG, Jilaihawi H, Virmani R, Anwaruddin S, Thourani VH, Nazif T, Mangner N, Woitek F, Krishnaswamy A, Mick S, Chakravarty T, Nakamura M, McCabe JM, Satler L, Zajarias A, Szeto WY, Svensson L, Alu MC, White RM, Kraemer C, Parhizgar A, Leon MB, Linke A; SENTINEL Trial Investigators. Protection Against Cerebral Embolism During Transcatheter Aortic Valve Replacement. J Am Coll Cardiol 2017;69:367-77
  • Ndunda PM, Vindhyal MR, Muutu TM, Fanari Z. Clinical Outcomes of Sentinel Cerebral Protection System Use During Transcatheter Aortic Valve Replacement: A Systematic Review and Meta-Analysis. Cardiovasc Revasc Med 2020;21:717-22
  • Kapadia SR, Makkar R, Leon M, Abdel-Wahab M, Waggoner T, Massberg S, Rottbauer W, Horr S, Sondergaard L, Karha J, Gooley R, Satler L, Stoler RC, Messé SR, Baron SJ, Seeger J, Kodali S, Krishnaswamy A, Thourani VH, Harrington K, Pocock S, Modolo R, Allocco DJ, Meredith IT, Linke A; PROTECTED TAVR Investigators. Cerebral Embolic Protection during Transcatheter Aortic-Valve Replacement. N Engl J Med 2022;387:1253-63
  • Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, Savovic J, Schulz KF, Weeks L, Sterne JA; Cochrane Bias Methods Group; Cochrane Statistical Methods Group. The Cochrane Collaboration’s tool for assessing risk of bias in randomised trials. BMJ 2011;343:d5928
  • Sterne JA, Hernán MA, Reeves BC, Savović J, Berkman ND, Viswanathan M, Henry D, Altman DG, Ansari MT, Boutron I, Carpenter JR, Chan AW, Churchill R, Deeks JJ, Hróbjartsson A, Kirkham J, Jüni P, Loke YK, Pigott TD, Ramsay CR, Regidor D, Rothstein HR, Sandhu L, Santaguida PL, Schünemann HJ, Shea B, Shrier I, Tugwell P, Turner L, Valentine JC, Waddington H, Waters E, Wells GA, Whiting PF, Higgins JP. ROBINS-I: A tool for assessing risk of bias in non-randomised studies of interventions. BMJ 2016;355:i4919
  • Seeger J, Gonska B, Otto M, Rottbauer W, Wöhrle J. Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement Significantly Reduces Death and Stroke Compared With Unprotected Procedures. JACC Cardiovasc Interv 2017;10:2297-303
  • Wan X, Wang W, Liu J, Tong T. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol 2014;14:135
  • Harbord RM, Egger M, Sterne JA. A modified test for small-study effects in meta-analyses of controlled trials with binary endpoints. Stat Med 2006;25:3443-57
  • Viechtbauer W. Conducting Meta-Analyses in R with the metafor Package. J Stat Soft. [Internet]. 2010;
  • Nombela-Franco L, Webb JG, de Jaegere PP, Toggweiler S, Nuis RJ, Dager AE, Amat-Santos IJ, Cheung A, Ye J, Binder RK, van der Boon RM, Van Mieghem N, Benitez LM, Pérez S, Lopez J, San Roman JA, Doyle D, Delarochellière R, Urena M, Leipsic J, Dumont E, Rodés-Cabau J. Timing, predictive factors, and prognostic value of cerebrovascular events in a large cohort of patients undergoing transcatheter aortic valve implantation. Circulation 2012;126:3041-53
  • Miller DC, Blackstone EH, Mack MJ, Svensson LG, Kodali SK, Kapadia S, Rajeswaran J, Anderson WN, Moses JW, Tuzcu EM, Webb JG, Leon MB, Smith CR. Transcatheter (TAVR) versus surgical (AVR) aortic valve replacement: occurrence, hazard, risk factors, and consequences of neurologic events in the PARTNER trial. J Thorac Cardiovasc Surg 2012;143:832-43
  • Radwan Y, Al-Abcha A, Salam MF, Khor SY, Prasad RM, Elshafie A, Abela G. Meta-analysis of the Safety and Efficacy of the Sentinel Cerebral Protection System in Transcatheter Aortic Valve Implantation. Am J Cardiol 2021;152:169-70
  • Eggebrecht H, Schmermund A, Voigtländer T, Kahlert P, Erbel R, Mehta RH. Risk of stroke after transcatheter aortic valve implantation (TAVI): a meta-analysis of 10,037 published patients. EuroIntervention 2012;8:129-38
  • Hatfield SA, Nores MA, James TM, Rothenberg M, Kapila A, Cubeddu RJ, Stamou SC. Predictors and outcomes of stroke after transcatheter aortic valve replacement. J Card Surg 2020;35:21-7
  • Kahlert P, Knipp SC, Schlamann M, Thielmann M, Al-Rashid F, Weber M, Johansson U, Wendt D, Jakob HG, Forsting M, Sack S, Erbel R, Eggebrecht H. Silent and apparent cerebral ischemia after percutaneous transfemoral aortic valve implantation: a diffusion-weighted magnetic resonance imaging study. Circulation 2010;121:870-8
  • Vermeer SE, Prins ND, den Heijer T, Hofman A, Koudstaal PJ, Breteler MM. Silent brain infarcts and the risk of dementia and cognitive decline. N Engl J Med 2003;348:1215-22
  • De Carlo M, Liga R, Migaleddu G, Scatturin M, Spaccarotella C, Fiorina C, Orlandi G, De Caro F, Rossi ML, Chieffo A, Indolfi C, Reimers B, Cosottini M, Petronio AS. Evolution, Predictors, and Neurocognitive Effects of Silent Cerebral Embolism During Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv 2020;13:1291-300
  • Stachon P, Kaier K, Heidt T, Wolf D, Duerschmied D, Staudacher D, Zehender M, Bode C, von Zur Mühlen C. The Use and Outcomes of Cerebral Protection Devices for Patients Undergoing Transfemoral Transcatheter Aortic Valve Replacement in Clinical Practice. JACC Cardiovasc Interv 2021;14:161-8
  • Alkhouli M, Alqahtani F, Harris AH, Hohmann SF, Rihal CS. Early Experience With Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement in the United States. JAMA Intern Med 2020;180:783-4
  • Butala NM, Makkar R, Secemsky EA, Gallup D, Marquis-Gravel G, Kosinski AS, Vemulapalli S, Valle JA, Bradley SM, Chakravarty T, Yeh RW, Cohen DJ. Cerebral Embolic Protection and Outcomes of Transcatheter Aortic Valve Replacement: Results From the Transcatheter Valve Therapy Registry. Circulation 2021;143:2229-40
  • Kroon HG, van der Werf HW, Hoeks SE, van Gils L, van den Berge FR, El Faquir N, Rahhab Z, Daemen J, Poelman J, Schurer RAJ, van den Heuvel A, de Jaegere P, van der Harst P, Van Mieghem NM. Early Clinical Impact of Cerebral Embolic Protection in Patients Undergoing Transcatheter Aortic Valve Replacement. Circ Cardiovasc Interv 2019;12:e007605
  • Khan MZ, Zahid S, Khan MU, Kichloo A, Ullah W, Sattar Y, Munir MB, Singla A, Goldsweig AM, Balla S. Use and outcomes of cerebral embolic protection for transcatheter aortic valve replacement: A US nationwide study. Catheter Cardiovasc Interv 2021;98:959-68
  • Seeger J, Kapadia SR, Kodali S, Linke A, Wöhrle J, Haussig S, Makkar R, Mehran R, Rottbauer W, Leon M. Rate of peri-procedural stroke observed with cerebral embolic protection during transcatheter aortic valve replacement: a patient-level propensity-matched analysis. Eur Heart J 2019;40:1334-40
  • Donà C, Koschutnik M, Nitsche C, Winter MP, Seidl V, Siller-Matula J, Mach M, Andreas M, Bartko P, Kammerlander AA, Goliasch G, Lang I, Hengstenberg C, Mascherbauer J. Cerebral Protection in TAVR-Can We Do Without? A Real-World All-Comer Intention-to-Treat Study-Impact on Stroke Rate, Length of Hospital Stay, and Twelve-Month Mortality. J Pers Med 2022;12:320
  • Isogai T, Vanguru HR, Krishnaswamy A, Agrawal A, Spilias N, Shekhar S, Saad AM, Verma BR, Puri R, Reed GW, Popović ZB, Unai S, Yun JJ, Uchino K, Kapadia SR. Cerebral embolic protection and severity of stroke following transcatheter aortic valve replacement. Catheter Cardiovasc Interv 2022;100:810-20
  • Seeger J, Falk V, Hildick-Smith DJ, Bleiziffer S, Blackman DJ, Abdel-Wahab M, Allocco DJ, Meredith IT, Wöhrle J, Van Mieghem NM. Insights on Embolic Protection, Repositioning, and Stroke: A Subanalysis of the RESPOND Study. J Interv Cardiol 2020;2020:3070427
  • Voss S, Deutsch MA, Schechtl J, Erlebach M, Sideris K, Lange R, Bleiziffer S. Impact of a Two-Filter Cerebral Embolic Protection Device on the Complexity and Risk of Transcatheter Aortic Valve Replacement. Thorac Cardiovasc Surg 2020;68:616-22

Volume 10 – Number 1

View full issue

Download this article
Keywords
  • aortic stenosis
  • degenerative valve
  • TAVI
Authors
  • Gao Fei
  • Jonathan Yap
  • Kay Woon Ho
  • Mohammed Rizwan Amanullah
  • Nicholas Tan
  • See Hooi Ewe
  • Sivaraj Govindasamy
  • Soo Teik Lim
  • Victor Tar Toong Chao
  • Zameer Abdul Aziz
AsiaIntervention
  • Readers
    • Archives
    • Subscribe to the newsletter
    • Contact us
  • About the journal
    • Overview
    • Editorial Board
    • Masthead
  • Services
    • Advertising in AsiaIntervention
    • Article reprints
    • Publication calendar
    • Rights & Permissions
  • Authors
    • Authors guidelines
    • Submit your paper
  • Legal
    • Disclaimer
    • Cookies Policy
    • Privacy Policy
    • Legal Notice
  • Follow us
    • Facebook
    • X
    • LinkedIn
Online ISSN 2491-0929 - Print ISSN 2426-3958
© 2015-2025 Europa Group - All rights reserved